Dose Comparison Between Gafchromic Film, XiO, And Monaco Treatment Planning Systems In A Novel Pelvic Phantom That Contains A Titanium Hip Prosthesis

Nicholas Ade, F.C.P. du Plessis
Published 2017 · Medicine
Cite This
Download PDF
Analyze on Scholarcy
Share
Abstract The presence of metallic prostheses during external beam radiotherapy of malignancies in the pelvic region has the potential to strongly influence the dose distribution to the target and to tissue surrounded by the prostheses. This study systematically investigates the perturbation effects of unilateral titanium prosthesis on 6 and 15 MV photon beam dose distributions using Gafchromic EBT2 film measurements in a novel pelvic phantom made out of a stack of nylon slices. Comparisons were also made between the film data and dose calculations made on XiO and Monaco treatment planning systems. The collapsed cone algorithm was chosen for the XiO and the Monte Carlo algorithm used on Monaco is XVMC. Transmission measurements were taken using a narrow‐beam geometry to determine the mass attenuation coefficient of nylon = 0.0458 cm2/g and for a water‐equivalent RW3 phantom, it was 0.0465 cm2/g. The perturbation effects of the prosthesis on dose distributions were investigated by measuring and comparing dose maps and profiles. The magnitude of dose perturbations was quantified by calculating dose enhancement and reduction factors using field sizes of 3 × 3, 5 × 5, 10 × 10, and 15 × 15 cm2. For the studied beams and field sizes, dose enhancements between 21 and 30% and dose reductions between 15 and 21% were observed at the nylon‐prosthesis interface on the proximal and distal sides of the prosthesis for film measurements. The dose escalation increases with beam energy, and the dose reduction due to attenuation decreases with increasing beam energy when compared to unattenuated beam data. A comparison of film and XiO depth doses for the studied fields gave relative errors between 1.1 and 23.2% at the proximal and distal interfaces of the Ti prosthesis. Also, relative errors < 4.0% were obtained between film and Monaco dose data outside the prosthesis for 6 and 15 MV lateral opposing fields.
This paper references
10.1118/1.3147204
The accuracy of the pencil beam convolution and anisotropic analytical algorithms in predicting the dose effects due to attenuation from immobilization devices and large air gaps.
A Gray (2009)
10.1016/0360-3016(88)90279-9
Effect of a femoral head prosthesis on megavoltage beam radiotherapy.
Peter J. Biggs (1988)
10.1088/1742-6596/74/1/021016
Characterization of materials for prosthetic implants using the BEAMnrc Monte Carlo code
E. Spezi (2007)
American Association of Physics in Medicine Radiation Therapy. Task Group 21: a protocol for the determination of absorbed dose from high-energy photon and electron beams
(1983)
10.1118/1.595446
A protocol for the determination of absorbed dose from high-energy photon and electron beams.
Aapm (1983)
10.1016/j.radmeas.2008.01.019
Evaluation of the water equivalence of solid phantoms using gamma ray transmission measurements
Robin E. Hill (2008)
10.1118/1.1592641
Tissue-equivalent materials for construction of tomographic dosimetry phantoms in pediatric radiology.
A. Kyle Jones (2003)
10.1118/1.1797531
Dose enhancement close to platinum implants for the 4, 6, and 10 MV stereotactic radiosurgery.
Joel Y C Cheung (2004)
Task Group 21 : a protocol for the determination of absorbed dose from high - energy photon and electron beams
MF Chan (1983)
10.1120/jacmp.v12i2.3295
Dose perturbation in the radiotherapy of breast cancer patients implanted with the Magna‐Site: a Monte Carlo study
Christos M Chatzigiannis (2011)
10.1088/0031-9155/48/20/003
Dose perturbation in the presence of metallic implants: treatment planning system versus Monte Carlo simulations.
Elinore Wieslander (2003)
Kn€ o€ os T. Dose perturbation in the presence of metallic implants: treatment planning system versus Monte Carlo simulations
E Wieslander (2003)
10.1118/1.4754797
An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.
David Lewis (2012)
Precise radiochromic film dosimetry using a flat-bed document scanner.
Slobodan Devic (2005)
10.4236/ijmpcero.2014.31005
Is It Possible to Publish a Calibration Function for Radiochromic Film?
Maria F. Chan (2014)
10.1016/S0360-3016(01)02592-5
A study on beams passing through hip prosthesis for pelvic radiation treatment.
George X. Ding (2001)
10.1120/jacmp.v11i1.3030
Acrylonitrile Butadiene Styrene (ABS) plastic‐based low cost tissue equivalent phantom for verification dosimetry in IMRT
Rajesh Kumar (2009)
10.1002/047120918X.emb0953
Monte Carlo Calculations
Arieh Warshel (2002)
10.1088/0031-9155/52/14/013
Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance.
Martina Fuss (2007)
10.1016/S0958-3947(02)00245-5
Radiotherapy dose calculations in the presence of hip prostheses.
Paul J. Keall (2003)
10.1118/1.1565113
Dosimetric considerations for patients with HIP prostheses undergoing pelvic irradiation. Report of the AAPM Radiation Therapy Committee Task Group 63.
Chester Stanley Reft (2003)
10.1007/s11604-007-0181-z
Dose attenuation effect of hip prostheses in a 9-MV photon beam: commercial treatment planning system versus Monte Carlo calculations
Asghar Mesbahi (2007)
10.1016/j.meddos.2005.06.001
A case study of radiotherapy planning for a bilateral metal hip prosthesis prostate cancer patient.
Andy W. Su (2005)
10.1118/1.2362876
Clinical use of EBT model Gafchromic™ film in radiotherapy.
Christian Fiandra (2006)
10.1118/1.596345
Backscatter dose perturbation at high atomic number interfaces in megavoltage photon beams.
Indra J. Das (1989)
10.1120/jacmp.v13i1.3347
An opposed matched field IMRT technique for prostate cancer patients with bilateral prosthetic hips
Saeid Fattahi (2012)
10.3906/FIZ-1401-15
Investigation of the Effects of Plaster on Radiation Doses Given to Patients
Onur Karaman (2015)
10.4103/2228-7477.83522
Review of Fast Monte Carlo Codes for Dose Calculation in Radiation Therapy Treatment Planning
Keyvan Jabbari (2011)
10.1016/0360-3016(90)90115-Z
Influence of hip prostheses on high energy photon dose distributions.
Claudio H. Sibata (1990)
10.1016/S0958-3947(01)00079-6
Intensity-modulated radiotherapy for a prostate patient with a metal prosthesis.
Jong Hyun Kung (2001)
10.1016/0360-3016(91)90210-U
Planning of radiotherapy for patients with hip prosthesis.
Martin Erlanson (1991)
10.1007/s13246-011-0102-4
Determination of RW3-to-water mass-energy absorption coefficient ratio for absolute dosimetry
Katrina Y. T. Seet (2011)
10.1016/S0958-3947(99)00021-7
Dose attenuation through a titanium alloy hip prosthesis.
Tony Y. Eng (2000)
10.1088/0031-9155/43/12/009
Dose perturbation caused by high-density inhomogeneities in small beams in stereotactic radiosurgery.
Surendra N. Rustgi (1998)
Clinical use of EBT model Gafchromic film in radiotherapy.
Christian Fiandra (2006)
10.1046/j.1440-1673.2000.00816.x
Effect of hip prostheses on radiotherapy dose.
Martin G Carolan (2000)
B€ ol€ ukdem_Ir mh. The effects of plaster on radiation doses given to patients
O Karaman (2015)
10.1120/jacmp.v14i2.4111
Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification
Valeria Casanova Borca (2013)



This paper is referenced by
Semantic Scholar Logo Some data provided by SemanticScholar