In Vivo Corrosion Of Modular Hip Prosthesis Components In Mixed And Similar Metal Combinations. The Effect Of Crevice, Stress, Motion, And Alloy Coupling.

Jeremy L Gilbert, Christine A. Buckley, Joshua J. Jacobs
Published 1993 · Materials Science, Medicine
Cite This
Download PDF
Analyze on Scholarcy
Share
One hundred forty-eight retrieved modular hip prostheses of both mixed (Ti-6Al-4V/Co-Cr) and similar (Co-Cr/Co-Cr) metal combinations were examined and positive evidence of corrosive attack was found in the conical taper region between head and stem. Significant corrosion was observed in both mixed and similar metal combinations with 16% of necks and 35% of heads (for mixed-metal cases), and 14% of necks and 23% of heads (for similar-metal cases) showing moderate to severe corrosive attack. There was a significant correlation between the percentage of prostheses with moderate to severe corrosion and the duration of implantation for both mixed and similar metal cases, indicating that this corrosion process is progressive in time. Moderate to severe corrosion was seen as early as 2.5 and 11 months (mixed and similar metals, respectively). Scanning electron microscopy and x-ray analysis identified several forms of corrosive attack in the cobalt-based component of the taper. These included, for both mixed and same metal combinations: preferential dissolution of cobalt, fretting, and pitting; mixed metals only: the formation of a Ti-Cr-Mo interfacial phase and interdendritic corrosion; and for similar metals: intergranular attack adjacent to grain boundaries enriched in molybdenum and silicon. It is hypothesized that the restricted crevice environment, coupled with high cyclic stresses which cause repeated fracture of the passive oxide films in the taper, result in an unstable electrochemical environment within the crevice for both the cobalt alloy and Ti-alloy passive films. The passivity of these alloys is subsequently lost and active attack of the taper results. Also, the repeated fracturing of the passive films will result in large amounts of corrosion products being formed. This corrosion and particulate accumulation could result in loss of mechanical integrity of the implants in vivo, create particles for third body wear, and release particles into the surrounding tissues.



This paper is referenced by
10.5006/2632
Technical Note: Is Corrosion a Threat to the Strength of the Taper Connection in Femoral Components of Total Hip Replacements?
Genymphas B. Higgs (2017)
10.1109/BIBE.2007.4375615
Retrieval Analysis of a Cementless Modular Total Hip Arthroplasty Prosthesis
Manish Paliwal (2007)
10.1097/CORR.0000000000000369
Do Well-functioning THAs Retrieved at Autopsy Exhibit Evidence of Fretting and Corrosion?
Jeffrey K. Lange (2018)
10.1520/STP159120140119
Studies of Modular Connections for Surgical Implant Devices
Jack E. Lemons (2015)
10.1053/J.SART.2013.01.013
Biologic Implications of Taper Corrosion in Total Hip Arthroplasty
H John Cooper (2012)
10.1002/jbm.b.31984
The cathodic behavior of titanium: Serum effect.
Francesco Contu (2012)
Wear and corrosion resistance of tantalum coating on titanium alloys for biomedical implant applications
Ay Ching Hee (2017)
10.1002/jbm.b.31943
In vivo oxide-induced stress corrosion cracking of Ti-6Al-4V in a neck-stem modular taper: Emergent behavior in a new mechanism of in vivo corrosion.
Jeremy L Gilbert (2012)
Tribology and Corrosion in CoCrMo Alloys and Similar Systems
Emily Hoffman (2017)
10.1243/09544119JEIM346
Femoral stem wear in cemented total hip replacement
H-Y Zhang (2008)
10.1177/0954411916648717
Assembly force and taper angle difference influence the relative motion at the stem–neck interface of bi-modular hip prostheses
Henning Haschke (2016)
10.1002/jor.23813
The onset of fretting at the head‐stem connection in hip arthroplasty is affected by head material and trunnion design under simulated corrosion conditions
Fiachra E. Rowan (2018)
10.1016/j.arth.2013.12.034
Mechanically assisted taper corrosion in modular TKA.
Christina M. Arnholt (2014)
10.2106/JBJS.M.01199
The relation between titanium taper corrosion and cobalt-chromium bearing wear in large-head metal-on-metal total hip prostheses: a retrieval study.
Florian Witt (2014)
10.1155/2015/758123
Head-Neck Taper Corrosion in Hip Arthroplasty
Saleem M Hussenbocus (2015)
10.1007/s11420-017-9566-4
Ceramic Bearings with Titanium Adapter Sleeves Implanted During Revision Hip Arthroplasty Show Minimal Fretting or Corrosion: a Retrieval Analysis
Chelsea N. Koch (2017)
10.1016/j.arth.2017.03.012
Metal-on-Metal Compared With Metal-on-Polyethylene: The Effect on Trunnion Corrosion in Total Hip Arthroplasty.
Brent A. Lanting (2017)
10.1016/j.medengphy.2013.11.009
Design parameters and the material coupling are decisive for the micromotion magnitude at the stem-neck interface of bi-modular hip implants.
Sabrina Jauch (2014)
10.1016/j.arth.2013.01.018
Taper corrosion in modular hip prostheses: analysis of serum metal ions in 19 patients.
Bart J Vundelinckx (2013)
10.1016/B978-0-08-055294-1.00013-1
Electrochemical Behavior of Metals in the Biological Milieu
Jeremy L Gilbert (2011)
10.1002/jbm.b.31861
Electrochemical investigation of chromium oxide-coated Ti-6Al-4V and Co-Cr-Mo alloy substrates.
Viswanathan Swaminathan (2011)
10.1016/J.CORSCI.2010.09.067
The voltage-dependent electrochemical impedance spectroscopy of CoCrMo medical alloy using time-domain techniques: Generalized Cauchy-Lorentz, and KWW-Randles functions describing non-ideal interfacial behaviour
Morteza Haeri (2011)
10.1016/J.WEAR.2017.08.011
The effect of simulated inflammatory conditions and pH on fretting corrosion of CoCrMo alloy surfaces
Yangping Liu (2017)
10.1016/j.clinbiomech.2017.04.009
Influence of the compliance of a patient's body on the head taper fixation strength of modular hip implants
Annika Krull (2017)
10.1002/jbm.b.33734
The biological response to orthopaedic implants for joint replacement: Part I: Metals.
Emmanuel Gibon (2017)
10.1007/s11999-017-5486-3
Alloy Microstructure Dictates Corrosion Modes in THA Modular Junctions
Robin Pourzal (2017)
10.1155/2016/1695906
Fretting and Corrosion in Modular Shoulder Arthroplasty: A Retrieval Analysis
Johannes A Eckert (2016)
10.1002/jbm.b.33499
The effect of fretting associated periodic cathodic potential shifts on the electrochemistry and in vitro biocompatibility of commercially pure titanium.
Alexandra Adrienne Ciolko (2016)
10.1520/STP159120140122
Fretting Corrosion and Modularity: A Critical Review of the Literature and Three Registries
John M. Cuckler (2015)
10.19080/JOJOOS.2018.12.555860
Using Large Femoral Head as A Bearing Surface in Total Hip Arthroplasty
Myung Rae Cho (2018)
10.1016/j.arth.2020.01.016
What Surgeons Need to Know About Adverse Local Tissue Reaction in Total Hip Arthroplasty.
Deborah J Hall (2020)
Validation of an Optical CMM for the Measurement of Wear at the Taper Interface in Total Hip Replacement
(2016)
See more
Semantic Scholar Logo Some data provided by SemanticScholar