The measurement of absorbed dose to water in a solid-phantom may require a conversion factor because it may not be radiologically equivalent to water. One phantom developed for the use of dosimetry is a solid water, RW3 white-polystyrene material by IBA. This has a lower mass-energy absorption coefficient than water due to high bremsstrahlung yield, which affects the accuracy of absolute dosimetry measurements. In this paper, we demonstrate the calculation of mass-energy absorption coefficient ratios, relative to water, from measurements in plastic water and RW3 with an Elekta Synergy linear accelerator (6 and 10 MV photon beams) as well as Monte Carlo modeling in BEAMnrc and DOSXYZnrc. From this, the solid-phantom-to-water correction factor was determined for plastic water and RW3.
This paper references
10.1118/1.597652
White polystyrene as a substitute for water in high energy photon dosimetry.
Gunter Christ (1995)
Monte Carlo Techniques in Radiotherapy
David William Oliver Rogers (2002)
10.1016/j.radmeas.2008.01.019
Evaluation of the water equivalence of solid phantoms using gamma ray transmission measurements
Robin E. Hill (2008)
10.1088/0031-9155/45/4/313
Converting absorbed dose to medium to absorbed dose to water for Monte Carlo based photon beam dose calculations
Jeffrey V. Siebers (2000)
10.1007/s12194-010-0082-1
Perturbation correction factors for cylindrical ionization chambers in high-energy electron beams
Takeshi Ono (2010)
10.1088/0031-9155/51/13/R17
Fifty years of Monte Carlo simulations for medical physics.
David William Oliver Rogers (2006)
10.1118/1.2012807
Absorbed dose to water reference dosimetry using solid phantoms in the context of absorbed-dose protocols.
Jan Seuntjens (2005)
10.1088/0031-9155/39/10/004
Absorbed-dose calibrations in high-energy photon beams at the National Physical Laboratory: conversion procedure.
Jean Ellen Burns (1994)
10.6028/nist.ir.5632
Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest
John H. Hubbell (1995)
AAA photon dose calculation in Eclipse, Varian documentation RAD #7170B
J Sievinen (2005)
10.1118/1.3151809
Monte Carlo calculations of correction factors for plastic phantoms in clinical photon and electron beam dosimetry.
Fujio Araki (2009)



This paper is referenced by
Semantic Scholar Logo Some data provided by SemanticScholar